Sanitary Temperature Transmitter

The ADDA-T-GW sanitary temperature transmitter is designed for temperature measurement in food, beverage, pharmaceutical, and bioengineering applications. It uses an all-stainless-steel structure, 355° rotating LCD module, sanitary dead-angle-free design, HART protocol support, 4–20 mA DC + HART or Modbus-RTU/RS485 output, and IP67 protection.

Authoritative Qualification Certifications

We hold ISO quality certification, national high-tech enterprise qualification, China measuring instrument type approval, and explosion-proof and safety certifications to support reliable product quality.

Custom Engineering

Configurable materials, connections, signals, displays, and installation solutions for different process environments.

Product Features

The ADDA-T-GW Series sanitary temperature transmitter uses RTD sensing for food, beverage, pharmaceutical, and bioengineering applications. It features an all-stainless-steel structure and a dot-matrix LCD module that supports 355° rotation. The hygienic design has no dead spaces and is designed to comply with FDA requirements. Key features include:

  • Sealed with silicone rubber or epoxy resin for vibration and moisture resistance; suitable for harsh operating environments.
  • High accuracy, low quiescent power consumption, stable and reliable operation, and long service life.
  • Hygienic design complies with EHEDG, 3A, FDA, and GMP requirements.
  • Wetted material surface roughness down to Ra 0.4.
  • Supports HART communication for local or remote parameter adjustment and monitoring through a handheld communicator.

Applications

Temperature measurement in food, beverage, pharmaceutical, and other sanitary applications.

Measuring Medium

Media compatible with the wetted materials.

 

Performance Specifications

Accuracy±0.1%, ±0.2%, 0.5% Upper Range Limit,  see specifications for details
Range-50~400 degrees C, see selection table for details
Annual Stability±0.05% Span/year
Voltage InfluenceWhen the supply voltage varies within 10.5V/16.5V~36V DC, the changes in zero point and span shall not exceed ±0.001%URL/V
Output Signal4mA~20mADC+HART

Modbus-RTU/RS485

Protection GradeIP67

Range

Sensor TypeMeasuring Range degrees CMinimum Measurable Range degrees CMeasurement Accuracy (whichever is greater) degrees C
Pt100-50~40050±0.2% range or ±0.2
Pt1000-50~35050±0.2% range or ±0.2

Environmental Conditions

ItemCondition
Operating Ambient Temperature Range-40 degrees C ~ 85 degrees C, LCD display: -20 degrees C ~ 70 degrees C
Storage Ambient Temperature Range-40 degrees C ~ 100 degrees C, LCD display: -40 degrees C ~ 85 degrees C
Operating Ambient Humidity Range5%RH~100%RH@40 degrees C
Hazardous AreaPCECCertificate No.: CE19.2398X Certificate No.: CE19.3332

Ex ia IIC T6 Ga Ex d IIC T6 Gb

Electromagnetic Compatibility Environment

No.Test ItemBasic StandardTest ConditionPerformance Level
1Radiated Disturbance (Enclosure)GB/T 9254.1/CISPR2230MHz-1000MHzB (Note 2)
2Conducted Disturbance (DC Power Port)GB/T 9254.1/CISPR220.15MHz-30MHzB (Note 2)
3Electrostatic Discharge (ESD) ImmunityGB/T 17626.2/IEC61000-4-24kV (contact), 8kV (air)A (Note 1)
4RF Electromagnetic Field ImmunityGB/T 17626.3/IEC61000-4-310V/m (80MHz-1GHz)A (Note 1)
5Power Frequency Magnetic Field ImmunityGB/T 17626.8/IEC61000-4-830A/mA (Note 1)
6Electrical Fast Transient/Burst ImmunityGB/T 17626.4/IEC61000-4-42kV (5/50ns, 100kHz)A (Note 1)
7Surge ImmunityGB/T 17626.5/IEC61000-4-54kV (line-to-line)

2kV (line-to-ground)

A (Note 1)
8Conducted Disturbance Immunity Induced by RF FieldsGB/T 17626.6/IEC61000-4-63V (150kHz-80MHz)A (Note 1)
Note 1: For performance level A, performance remains normal within the technical specification limits

Note 2: For performance level B, function or performance may temporarily decrease or be lost, but can recover automatically, actual operating status, storage and data remain unchanged

Power Supply and Load Conditions

ItemOperating Condition 
Power Supply VoltageHART communication protocol: 16.5V~36V DC
Intrinsically safe version: HART communication protocol: 18.5V~28V DC
Modbus-RTU/RS485: 12V~36V DC
Load Resistance0Ω~2119Ω during operation

250Ω~600Ω for HART communication

Transmission Distance< 1000m
Power Consumption
4mA~20mA<= 500mW@24V DC, 20.8mA
Modbus-RTU/RS485<= 240mW@24V DC, 10mA

Related Products

Case Showcase

Frequently Asked Questions

Find quick answers about product selection, customization, technical parameters, delivery, and support for ADDA industrial measurement instruments.

官网询盘

Please provide the measured medium, measuring range, process temperature, process pressure, connection type, output signal, power supply, installation environment, and communication method if remote monitoring is required.

Yes. ADDA can provide customized solutions based on measuring range, process connection, wetted material, display type, communication protocol, installation method, and working environment.

Many ADDA instruments are available with explosion-proof or intrinsically safe configurations. Please confirm the site classification and certification requirements before selection.

Yes. Our team can help recommend a suitable instrument configuration after reviewing your application conditions, including medium, range, connection, output signal, and installation requirements.

Lead time depends on product type, quantity, customization requirements, and certification needs. Standard products usually have a shorter delivery time, while customized models may require additional production time.

Contact Us for Cooperation

官网询盘